How can I check the MFFT is operating at the correct temperature?

Simply place the instrument on range one, allow the platen to stabilise and take note of the temperatures between point 1 and 10. Point 1 should read -5°C and point 10 should read 13°C There should be an even gradient between all points.

Related products

Aesthetix®: The Ultimate Surface Finish Solution​

  • Fully characterise all aspects of appearance perception - Why does this matter?
  • Modular solution: Surface Brilliance | Cross-cut Adhesion | Texture | Effect Finish | Polishing Quality
  • Aesthetix® can be used in-laboratory, portable in-field, inline in the production environment or as part of a Cobot / Robot solution
  • Measures flat surfaces, curved parts and small surface areas

Flow cup afnor

AFNOR Viscosity Flow Cup

  • 2mm - 8mm orifice diameter
  • 15 – 3000 cSt viscosity range
  • 25 -150 sec flow times
  • Supplied with a certificate of conformity
  • Sheen 419/2.5, 419/4, 419/6 & 419/8 equivalent

Flow cup astm iso

ASTM D1200 Viscosity Flow Cup

  • 2.1mm - 5.8mm orifice diameter
  • 10 – 1200 cSt viscosity range
  • 30-100 sec flow times
  • Supplied with a certificate of conformity
  • Sheen 406/1, 406/2, 406/3, 406/4 & 406/5 equivalent

FAQ: How can I check the MFFT is operating at the correct temperature?

Rhopoint Americas
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.