+1-888-214-6090

How can I check the MFFT is operating at the correct temperature?

Simply place the instrument on range one, allow the platen to stabilise and take note of the temperatures between point 1 and 10. Point 1 should read -5°C and point 10 should read 13°C There should be an even gradient between all points.

Related products

Flow cup afnor

AFNOR Viscosity Flow Cup

  • 2mm - 8mm orifice diameter
  • 15 – 3000 cSt viscosity range
  • 25 -150 sec flow times
  • Supplied with a certificate of conformity
  • Sheen 419/2.5, 419/4, 419/6 & 419/8 equivalent

Flow cup astm iso

ASTM D1200 Viscosity Flow Cup

  • 2.1mm - 5.8mm orifice diameter
  • 10 – 1200 cSt viscosity range
  • 30-100 sec flow times
  • Supplied with a certificate of conformity
  • Sheen 406/1, 406/2, 406/3, 406/4 & 406/5 equivalent

FAQ: How can I check the MFFT is operating at the correct temperature?