How can UKAS/ISO 17025 certification be maintained during service at Rhopoint Instruments GmbH and Rhopoint Americas Inc?

How can UKAS/ISO 17025 certification be maintained during service at Rhopoint Instruments GmbH and Rhopoint Americas Inc?

Calibration & Metrological Traceability

Rhopoint Instruments operates a central ISO/IEC 17025‑accredited gloss laboratory in the United Kingdom, UKAS laboratory number 32245, with regional calibration services in Germany and the United States.

During recalibration at a regional location, gloss instruments are adjusted to Rhopoint manufacturer traceable standards, and the reference standard is replaced with a new standard certified by the UK laboratory in accordance with ISO/IEC 17025.

This combination ensures metrological traceability to the accredited UK laboratory while maintaining manufacturer‑controlled calibration through local service.

The UK laboratory is accredited to ISO/IEC 17025 by UKAS for gloss calibration and the certification of high‑gloss reference tiles.

Neither Rhopoint Instruments GmbH nor Rhopoint Americas is itself ISO/IEC 17025‑accredited; regional calibration services are traceable to, but not performed within, the accredited laboratory.

Related products

FAQ: How can UKAS/ISO 17025 certification be maintained during service at Rhopoint Instruments GmbH and Rhopoint Americas Inc?

Rhopoint Americas
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.